Back to Library
Utilizing Flow Cytometry Within a Quality Control Process for an iPSC-Based Manufacturing Facility
-
Posters
Loading PDF document…
Download Utilizing Flow Cytometry Within a Quality Control Process for an iPSC-Based Manufacturing Facility PDF
Enter your details to unlock PDF access and receive your download link.
More information
The poster Utilizing Flow Cytometry Within a Quality Control Process for an iPSC-Based Manufacturing Facility presents Axol Bioscience’s approach to integrating flow cytometry into its ISO-compliant QC pipeline for iPSC-derived cell manufacturing.
Key highlights include:
- QC Workflow:
- Establish pass thresholds
- Assess cells via flow cytometry
- Analyze pass/fail status
- Report within QMS and Certificate of Analysis (CoA)
- Instrumentation:
- Guava® easyCyte and Attune NxT flow cytometers used for parallel validation
- Pluripotency Assessment:
- iPSCs tested for SSEA-1 (<10%), OCT3/4, SSEA-4, TRA-1-60, NANOG (>80%)
- Precursor Cell Validation:
- Monocytes (microglia precursors) tested for CD14, CD16, CD11b, CD206, CD163, SSEA-4
- Pass thresholds: >80% for lineage markers, <10% for SSEA-4
- Endpoint Cell Characterization:
- Cardiomyocytes (axo2508) tested for Cardiac Troponin T (>60%)
- Clear differentiation from iPSCs (1.62% expression)
- Consistency & Troubleshooting:
- Flow cytometry used to detect immature or aberrant batches
- Supports axoServices™ for external client QC
This QC framework ensures reproducibility, high-quality cell output, and project success across Axol’s manufacturing pipeline.
Featured Axol Products:
- axoCells™ Cardiomyocytes – ax2508
- axoCells™ Neural Stem Cells – ax0015